A cross-sectional study of the radiation dose and image quality of X-ray equipment used in IVR

Yohei Inaba, Koichi Chida, Ryota Kobayashi, Masayuki Zuguchi

Research output: Contribution to journalArticlepeer-review

21 Citations (Scopus)

Abstract

There are case reports of injuries caused by the radiation from interventional radiology (IVR) X-ray systems. Therefore, the management of radiation doses in IVR is important. However, no detailed report has evaluated image quality for a large number of IVR X-ray systems. As a result, it is unclear whether the image quality of the X-ray equipment currently used in IVR procedures is optimal. We compared the entrance surface doses and image quality of multiple IVR X-ray systems. This study was conducted in 2014 at 13 medical facilities using 18 IVR X-ray systems. We evaluated image quality and simultaneously measured the radiation dose. Entrance surface doses for fluoroscopy (duration, 1 min) and cineradiography (duration, 10 s) are measured using a 20-cm-thick acrylic plate and skin dose monitor. The image quality (such as spatial resolution and low-contrast detectability) of both fluoroscopy and cineradiography was evaluated using a QC phantom. For fluoroscopy, the average entrance surface dose using the 20-cm-thick acrylic plate was 13.9 (range 2.1-28.2) mGy/min. For cineradiography, the average entrance surface dose was 24.6 (range 5.1-49.3) mGy/10 s. We found positive correlations between radiation doses and image quality scores, in general, especially for fluoroscopy. The differences in surface dose among the 18 IVR X-ray systems were high (max/min, 9.7-fold for cineradiography; 13.4-fold for fluoroscopy). The differences in image quality scores (spatial resolution, low-contrast detectability, and dynamic range) were also very large. In general, there tended to be a correlation between radiation dose and image quality. Periodical measurements of the radiation dose and image quality of the X-ray equipment used for cineradiography and fluoroscopy in IVR are necessary. The need to minimize patient exposure requires that the dose be reduced to the minimum level that will generate an image with an acceptable degree of noise.

Original languageEnglish
Pages (from-to)391-401
Number of pages11
JournalJournal of applied clinical medical physics
Volume17
Issue number4
DOIs
Publication statusPublished - 2016

Keywords

  • Cineradiography
  • Fluoroscopy
  • Image quality
  • Interventional radiology (IVR)
  • Radiation dosimetry

ASJC Scopus subject areas

  • Radiation
  • Instrumentation
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'A cross-sectional study of the radiation dose and image quality of X-ray equipment used in IVR'. Together they form a unique fingerprint.

Cite this