Abstract
A new glass capillary microelectrode for L-glutamate is described using pulled glass capillaries (tip size, ∼12.5 μm) with a very small volume (-2 μl) of inner solution containing glutamate oxidase (GluOx) and ascorbate oxidase. The operation of the electrode is based on capillary action that samples L-glutamate into the inner solution. The enzyme reaction by GluOx generates hydrogen peroxide that is detected at an Os-gel-HRP polymer modified Pt electrode in a three-electrode configuration. The amperometric response behavior of the electrode was characterized in terms of the capillarity, response time, sensitivity and selectivity for measurements of L-glutamate. The currents at 0 V vs. Ag/AgCl increased linearly with the L-glutamate concentration from 10 to 150 μM for in vitro and in situ calibrations. The response was highly selective to L-glutamate over ascorbate, dopamine, serotonin and other amino acids. The detection of L-glutamate in the extracellular fluids of different regions of mouse hippocampal slices under stimulation of KCl was demonstrated.
Original language | English |
---|---|
Pages (from-to) | 55-60 |
Number of pages | 6 |
Journal | Analytical Sciences |
Volume | 19 |
Issue number | 1 |
DOIs | |
Publication status | Published - 2003 Jan 1 |