T1 - A novel support system for patient immobilization and transportation for daily computed tomographic localization of target prior to radiation therapy
AU - Nemoto, Kenji
AU - Seiji, Kazumasa
AU - Sasaki, Kazuya
AU - Kasamatsu, Nobutaka
AU - Fujishima, Toshie
AU - Ogawa, Yoshihiro
AU - Ariga, Hisanori
AU - Takeda, Ken
AU - Kimura, Tokihisa
AU - Yamada, Shogo
PY - 2003/3/15
Y1 - 2003/3/15
N2 - Purpose: To develop a method for quick and smooth transportation of patients from a computed tomography (CT) table to a linear accelerator (linac) table for confirming tumor center before radiation therapy. Methods and Materials: We developed a system using a subtable for patient immobilization that is transported via a customized stretcher. The patient lies on the subtable and is immobilized by a vacuum cushion and thermoplastic body cast. The subtable stretcher is used to carry the subtable from the CT table to the linac table. During transportation, the subtable is kept flat and shock to the subtable is carefully avoided. Between August 2001 and September 2002, a total of 9 patients with solitary upper lung tumors (superior to carina) were treated using this system. Results: Intrafractional tumor motion along the x (left-right), y (anterior-posterior), and z axis (superior-inferior) ranged from -2 mm to 2 mm, -2 mm to 2 mm, and -5 mm to 3 mm, respectively. The standard deviation of intrafractional tumor motion along the x, y, and z axis ranged from 0.5 mm to 1.5 mm, 0 mm to 1.7 mm, and 0.6 mm to 3.5 mm, respectively. Interfractional setup errors along the x, y, and z axis ranged from -5 mm to 4 mm, -6 mm to 8 mm, and -6 mm to 6 mm, respectively, and we could reduce interfractional setup errors in the majority of treatment sessions. Conclusion: We developed a system that allows patients to be immobilized and transported to verify tumor location on a daily basis. This system is highly useful for reducing setup errors.
AB - Purpose: To develop a method for quick and smooth transportation of patients from a computed tomography (CT) table to a linear accelerator (linac) table for confirming tumor center before radiation therapy. Methods and Materials: We developed a system using a subtable for patient immobilization that is transported via a customized stretcher. The patient lies on the subtable and is immobilized by a vacuum cushion and thermoplastic body cast. The subtable stretcher is used to carry the subtable from the CT table to the linac table. During transportation, the subtable is kept flat and shock to the subtable is carefully avoided. Between August 2001 and September 2002, a total of 9 patients with solitary upper lung tumors (superior to carina) were treated using this system. Results: Intrafractional tumor motion along the x (left-right), y (anterior-posterior), and z axis (superior-inferior) ranged from -2 mm to 2 mm, -2 mm to 2 mm, and -5 mm to 3 mm, respectively. The standard deviation of intrafractional tumor motion along the x, y, and z axis ranged from 0.5 mm to 1.5 mm, 0 mm to 1.7 mm, and 0.6 mm to 3.5 mm, respectively. Interfractional setup errors along the x, y, and z axis ranged from -5 mm to 4 mm, -6 mm to 8 mm, and -6 mm to 6 mm, respectively, and we could reduce interfractional setup errors in the majority of treatment sessions. Conclusion: We developed a system that allows patients to be immobilized and transported to verify tumor location on a daily basis. This system is highly useful for reducing setup errors.
KW - Computed tomography
KW - Daily localization
KW - Lung tumor
KW - Radiotherapy
UR - http://www.scopus.com/inward/record.url?scp=0242669955&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0242669955&partnerID=8YFLogxK
U2 - 10.1016/S0360-3016(02)04513-3
DO - 10.1016/S0360-3016(02)04513-3
M3 - Article
C2 - 12605990
AN - SCOPUS:0242669955
SN - 0360-3016
VL - 55
SP - 1102
EP - 1108
JO - International Journal of Radiation Oncology Biology Physics
JF - International Journal of Radiation Oncology Biology Physics
IS - 4
ER -