A position-sensitive scintillation detector for two-dimensional angular correlation of annihilation radiation using metal-package position-sensitive photomultiplier tubes

Koji Inoue, Yasuyoshi Nagai, Haruo Saito, Yasuyuki Nagashima, Toshio Hyodo, Shinichi Muramatsu, Shota Nagai

Research output: Contribution to journalArticlepeer-review

8 Citations (Scopus)

Abstract

We have constructed and tested a prototype of a new position sensitive γ-ray detector which consists of an array of 2.6×2.6×18 mm3 BGO scintillator blocks, a light guide, and four metal-package position-sensitive photomultiplier tubes (R5900-00-C8) recently developed by Hamamatsu Photonics Co. Ltd. Scalability of the detector of this type makes it possible to construct a larger detector using many PS-PMTs, which will be useful for the two-dimensional angular correlation of annihilation radiation apparatus.

Original languageEnglish
Pages (from-to)364-368
Number of pages5
JournalNuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Volume423
Issue number2-3
DOIs
Publication statusPublished - 1999 Mar 1