Abstract
An adaptive noise reduction process based on a gray-level concentrative filter is proposed for CW Doppler spectrogram. A graylevel concentrative filter as well as a convolution filter was tested to separate cardiac Doppler signals from the background noise in the histogram of the spectral amplitude as a function of time. By setting the cut-off threshold to the peak of the histogram of background noise amplitude after filtering, the background noise was significantly reduced without deteriorating the spectral signal. By setting the threshold to the valley between the histogram peaks of the background noise and spectral amplitudes after filtering, the maximum velocity was automatically traced in the spectrum as a function of time with a simple algorithm. A gray-level concentrative filter always showed a slightly better performance than a convolution filter with the same sliding window size in histogramically separating the Doppler signals from the background noise.
Original language | English |
---|---|
Pages (from-to) | 1453-1456 |
Number of pages | 4 |
Journal | Proceedings - IEEE Ultrasonics Symposium |
Volume | 2 |
Publication status | Published - 2004 Dec 1 |
Event | 2004 IEEE Ultrasonics Symposium - Montreal, Que., Canada Duration: 2004 Aug 23 → 2004 Aug 27 |