N2 - We have been studying the opening action of a monoleaflet tilting disc valve, (Bjork-Shiley Monostrut; sewing ring diameter: 23 mm, BSM), and reported last year that the valve opening motion depended on the early period flow rate pattern from 0 to 20 msec. This time, the Bjork-Shiley Spherical and the St. Vincent's valve were tested. All three valves are monoleaflet tilting disc valves with the same disc diameter but with disc shapes that are different from each other. Each valve was incorporated in a mitral position of a mock circulatory system driven by pneumatic pressure. Changes in valvular disc angle were measured at 500 frames per second using a high speed videocamera. We found the opening motion of the two valves depended on the early period flow rate pattern (from 0 to 20 msec; the same as the BSM valve). Using the moment of force values obtained, drag and lift coefficients (C(D) and C(L)) of each valve were calculated. Results showed that the C(D) and the C(L) were enlarged up to 3.0 x 104 at the early opening period. This result was speculated to be a cause of the influence of the flow pattern on the valve action at the same period.
AB - We have been studying the opening action of a monoleaflet tilting disc valve, (Bjork-Shiley Monostrut; sewing ring diameter: 23 mm, BSM), and reported last year that the valve opening motion depended on the early period flow rate pattern from 0 to 20 msec. This time, the Bjork-Shiley Spherical and the St. Vincent's valve were tested. All three valves are monoleaflet tilting disc valves with the same disc diameter but with disc shapes that are different from each other. Each valve was incorporated in a mitral position of a mock circulatory system driven by pneumatic pressure. Changes in valvular disc angle were measured at 500 frames per second using a high speed videocamera. We found the opening motion of the two valves depended on the early period flow rate pattern (from 0 to 20 msec; the same as the BSM valve). Using the moment of force values obtained, drag and lift coefficients (C(D) and C(L)) of each valve were calculated. Results showed that the C(D) and the C(L) were enlarged up to 3.0 x 104 at the early opening period. This result was speculated to be a cause of the influence of the flow pattern on the valve action at the same period.
KW - Drag coefficient
KW - High speed video camera
KW - Mechanical circulatory system
KW - Monoleaflet tilting disc valve
KW - Opening behavior
UR - http://www.scopus.com/inward/record.url?scp=0031918093&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0031918093&partnerID=8YFLogxK
M3 - Article
AN - SCOPUS:0031918093
SN - 0300-0818
VL - 27
SP - 158
EP - 163
JO - Japanese Journal of Artificial Organs
JF - Japanese Journal of Artificial Organs
IS - 1
ER -