TY - GEN
T1 - Complete erasing of ghost images caused by deeply trapped electrons on computed radiography plates
AU - Ohuchi, H.
AU - Kondo, Y.
PY - 2011
Y1 - 2011
N2 - The ghost images, i.e., latent image that is unerasable with visible light (LIunVL) and reappearing image appeared on computed radiography (CR) plates were completely erased by simultaneous exposing them to filtered ultraviolet light and visible light. Three different types of CR plates (Agfa, Kodak, and Fuji) were irradiated with 50 kV X-ray beams in the dose range 8.1 mGy to 8.0 Gy, and then conventionally erased for 2 h with visible light. The remaining LIunVL could be erased by repeating 6 h simultaneous exposures to filtered ultraviolet light and visible light. After the sixth round of exposure, all the LIunVL in the three types of CR plates were erased to the same level as in an unirradiated plate and no latent images reappeared after storage at 0°C for 14 days. The absorption spectra of deep centers were specified using polychromatic ultraviolet light from a deep-ultraviolet lamp. It was found that deep centers showed a dominant peak in the absorption spectra at around 324 nm for the Agfa and Kodak plates, and at around 320 nm for the Fuji plate, in each case followed by a few small peaks. After completely erasing CR plates, these peaks were no longer observed.
AB - The ghost images, i.e., latent image that is unerasable with visible light (LIunVL) and reappearing image appeared on computed radiography (CR) plates were completely erased by simultaneous exposing them to filtered ultraviolet light and visible light. Three different types of CR plates (Agfa, Kodak, and Fuji) were irradiated with 50 kV X-ray beams in the dose range 8.1 mGy to 8.0 Gy, and then conventionally erased for 2 h with visible light. The remaining LIunVL could be erased by repeating 6 h simultaneous exposures to filtered ultraviolet light and visible light. After the sixth round of exposure, all the LIunVL in the three types of CR plates were erased to the same level as in an unirradiated plate and no latent images reappeared after storage at 0°C for 14 days. The absorption spectra of deep centers were specified using polychromatic ultraviolet light from a deep-ultraviolet lamp. It was found that deep centers showed a dominant peak in the absorption spectra at around 324 nm for the Agfa and Kodak plates, and at around 320 nm for the Fuji plate, in each case followed by a few small peaks. After completely erasing CR plates, these peaks were no longer observed.
KW - X-ray
KW - complete erasing
KW - computed radiography (CR) plate
KW - deeply trapped electrons
KW - ghost image
KW - simultaneous exposure with filtered ultraviolet light and visible light
UR - http://www.scopus.com/inward/record.url?scp=79955785065&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=79955785065&partnerID=8YFLogxK
U2 - 10.1117/12.877970
DO - 10.1117/12.877970
M3 - Conference contribution
AN - SCOPUS:79955785065
SN - 9780819485038
T3 - Progress in Biomedical Optics and Imaging - Proceedings of SPIE
BT - Medical Imaging 2011
T2 - Medical Imaging 2011: Physics of Medical Imaging
Y2 - 13 February 2011 through 17 February 2011
ER -