Abstract
A small volume L glutamate online sensor was developed in order to monitor changes in the local concentration of L-glutamate released from cultured nerve cells. Syringe pump in the suction mode is used to sample extracellular fluid continuously from a glass micro-capillary and the concentration of L-glutamate can be determined by using a glassy carbon (GC) electrode modified with an Os-polyvinylpyridine mediator bottom film containing horseradish peroxidase and a bovine serum albumin top layer containing L-glutamate oxidase. The overall efficiency of L-glutamate detection with a sensor is 71% under optimum conditions due to an efficient enzymatic reaction at the modified electrode in the thin layer radial flow cell. As a result, we achieved a detection limit of 7-15 nM and a linear range of 50 nM to 10 μM. In an in vitro experiment, the extracellular fluid near a particular nerve cell can be sampled with this micro-pipet and continuously introduced into the modified GC electrode in the radial flow cell via suction provided by a syringe pump. The nerve cells are stimulated by KCl in a glass capillary and the L-glutamate concentration change can be monitored by changing the distance between the sampling piper and the nerve cells.
Original language | English |
---|---|
Pages (from-to) | 311-319 |
Number of pages | 9 |
Journal | Biosensors and Bioelectronics |
Volume | 12 |
Issue number | 4 |
DOIs | |
Publication status | Published - 1997 |
Keywords
- L-glutamate sensor
- capillary
- enzyme
- nerve cell
- online