Abstract
For the precise quantification of Ni-Pd-P alloys by inductively coupled plasma-optical emission spectrometry, non-spectral interferences of K should be corrected because the standard solution for P is usually prepared by using KH2PO4. An analytical method is suggested to correct the influence of K; after dissolution of the sample in HNO3-HCl, a standard solution of K is added to the sample solution so that the amount of K can be controlled to be the same as in the P standard solution with an error of ± 1 mg.
Original language | English |
---|---|
Pages (from-to) | 53-56 |
Number of pages | 4 |
Journal | BUNSEKI KAGAKU |
Volume | 53 |
Issue number | 1 |
DOIs | |
Publication status | Published - 2004 Jan |
Externally published | Yes |
Keywords
- HNO-HCl dissolution
- ICP-OES
- Influence of K
- Ni-Pd-P alloys
ASJC Scopus subject areas
- Analytical Chemistry