Abstract
Laser induced breakdown spectrometry (LIBS) combined with laser induced fluorescence spectrometry (LIFS) has been applied for detection of trace-level phosphorus in steel. The plasma induced by irradiation of Nd:YAG laser pulse for ablation was illuminated by the 3rd harmonic of Ti:Sapphire laser tuned to one of the resonant lines for phosphorus in the wavelength region of 253-256 nm. An excitation line for phosphorus was selected to give the highest signal-to-noise ratio. Fluorescence signals, P213.62 and P214.91 nm, were observed with high selectivity at the contents as low as several tens μg g- 1. Fluorescence intensities were in a good linear correlation with the contents. Fluorescence intensity ratio of a collisionally assisted line (213.62 nm) to a direct transition line (214.91 nm) was discussed in terms of the analytical conditions and experimental results were compared with a calculation based on rate equations. Since the fluorescence signal light in the wavelength range longer than 200 nm can be transmitted relatively easily, even through fiber optics of moderate length, LIBS/LIFS would be a versatile technique in on-site applications for the monitoring of phosphorus contents in steel.
Original language | English |
---|---|
Pages (from-to) | 884-890 |
Number of pages | 7 |
Journal | Spectrochimica Acta - Part B Atomic Spectroscopy |
Volume | 64 |
Issue number | 9 |
DOIs | |
Publication status | Published - 2009 Sept 1 |
Keywords
- LIBS
- Laser induced fluorescence
- Phosphorus in steel
ASJC Scopus subject areas
- Analytical Chemistry
- Atomic and Molecular Physics, and Optics
- Instrumentation
- Spectroscopy