TY - JOUR
T1 - Development of a dual-sided readout DOI-PET module using large-area monolithic mppc-arrays
AU - Kishimoto, Aya
AU - Kataoka, Jun
AU - Kato, Takuya
AU - Miura, Takamasa
AU - Nakamori, Takeshi
AU - Kamada, Kei
AU - Nakamura, Shigeyuki
AU - Sato, Kenichi
AU - Ishikawa, Yoshitaka
AU - Yamamura, Kazuhisa
AU - Kawabata, Nobuyuki
AU - Yamamoto, Seiichi
PY - 2013
Y1 - 2013
N2 - We are proposing a novel design for a module with depth of interaction (DOI) capability for gamma rays by measuring the pulse-height ratio of double-sided Multi-Pixel Photon Counters (MPPCs) coupled at both ends of a scintillation crystal block. Thanks to newly developed monolithic MPPC arrays consisting of 4 × 4 channels with a three-side buttable package, the module is very thin and compact, thereby enabling less dead space between each module when arranged into a fully designed gantry. To demonstrate our concept of a DOI measuring technique, we first made a 1-D crystal array consisting of five Ce-doped Gd3Al2Ga3O12 (Ce:GAGG) cubic crystals measuring 3× 3× 3\mm3 in size, separated by a layer of air approximately 10 μ m-thick. When the light signals output from both ends are read with the 3× 3 mm2 MPPCs, the position of each crystal is clearly distinguished. The same measurements were also made using Ce-doped (Lu,Y)2(SiO4)O(Ce:LYSO), achieving a similarly good separation. We then fabricated thin Ce:GAGG 2-D crystal arrays consisting of two types: [A] 4 × 4 matrix of 3× 3× 3mm} 3 pixels, and [B] 10 × 10 matrix of 0.8× 0.8× 5 mm3 pixels, with each pixel divided by a BaSO4 reflector 0.2 mm-thick. Then four arrays are laid on top of each other facing the DOI direction through a layer of air 10 μm-thick. We demonstrated that the 3-D position of each Ce:GAGG pixel is clearly distinguished in both the 2-D and DOI directions for type A and B when illuminated by 662 keV gamma rays. Average energy resolutions of 9.8 ± 0.8% and 11.8 ± 1.3% were obtained for types A and B, respectively. These results suggest that our proposed method is simple and offers promise in achieving both excellent spatial and energy resolutions for future medical imaging, particularly in positron emission tomography (PET).
AB - We are proposing a novel design for a module with depth of interaction (DOI) capability for gamma rays by measuring the pulse-height ratio of double-sided Multi-Pixel Photon Counters (MPPCs) coupled at both ends of a scintillation crystal block. Thanks to newly developed monolithic MPPC arrays consisting of 4 × 4 channels with a three-side buttable package, the module is very thin and compact, thereby enabling less dead space between each module when arranged into a fully designed gantry. To demonstrate our concept of a DOI measuring technique, we first made a 1-D crystal array consisting of five Ce-doped Gd3Al2Ga3O12 (Ce:GAGG) cubic crystals measuring 3× 3× 3\mm3 in size, separated by a layer of air approximately 10 μ m-thick. When the light signals output from both ends are read with the 3× 3 mm2 MPPCs, the position of each crystal is clearly distinguished. The same measurements were also made using Ce-doped (Lu,Y)2(SiO4)O(Ce:LYSO), achieving a similarly good separation. We then fabricated thin Ce:GAGG 2-D crystal arrays consisting of two types: [A] 4 × 4 matrix of 3× 3× 3mm} 3 pixels, and [B] 10 × 10 matrix of 0.8× 0.8× 5 mm3 pixels, with each pixel divided by a BaSO4 reflector 0.2 mm-thick. Then four arrays are laid on top of each other facing the DOI direction through a layer of air 10 μm-thick. We demonstrated that the 3-D position of each Ce:GAGG pixel is clearly distinguished in both the 2-D and DOI directions for type A and B when illuminated by 662 keV gamma rays. Average energy resolutions of 9.8 ± 0.8% and 11.8 ± 1.3% were obtained for types A and B, respectively. These results suggest that our proposed method is simple and offers promise in achieving both excellent spatial and energy resolutions for future medical imaging, particularly in positron emission tomography (PET).
KW - Depth of Interaction (DOI)
KW - Multi-Pixel Photon Counter (MPPC)
KW - Positron Emission Tomography (PET)
UR - http://www.scopus.com/inward/record.url?scp=84873628066&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84873628066&partnerID=8YFLogxK
U2 - 10.1109/TNS.2012.2233215
DO - 10.1109/TNS.2012.2233215
M3 - Article
AN - SCOPUS:84873628066
SN - 0018-9499
VL - 60
SP - 38
EP - 43
JO - IEEE Transactions on Nuclear Science
JF - IEEE Transactions on Nuclear Science
IS - 1
M1 - 6423841
ER -