Development of a multimodal ion-beam-analysis system for the 1-MV tandem Pelletron accelerator installed at Tohoku University

Kyoka Maruta, Sho Toyama, Misako Miwa, Hitoshi Takamura, Akihiro Ishii, Wataru Kada, Yohei Kikuchi, Shigeo Matsuyama

Research output: Contribution to journalArticlepeer-review

Abstract

To improve our ability to perform ion-beam analyses (IBA), we installed a new 1-MV 3SDH-2 tandem Pelletron accelerator at Tohoku University. It has two beam lines: a 15° line for heavy-ion-beam irradiation experiments and a 45° line for IBA experiments. For multimodal IBA experiments, we installed a newly designed and developed target chamber on the 45° beamline. We have since carried out a variety of fundamental and applied research projects using this facility. Here, we provide an overview of the facility, the target chamber developed for IBA, and some examples that involve the use of Rutherford backscattering spectroscopy, elastic-recoil detection analyses, particle-induced X-ray emission, and nuclear-reaction analysis techniques for specimens from different research fields.

Original languageEnglish
Article number165414
JournalNuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms
Volume554
DOIs
Publication statusPublished - 2024 Sept

Keywords

  • Compact accelerator
  • Multimodal IBA
  • PIXE
  • RBS
  • Y–O–H photochromic film

Fingerprint

Dive into the research topics of 'Development of a multimodal ion-beam-analysis system for the 1-MV tandem Pelletron accelerator installed at Tohoku University'. Together they form a unique fingerprint.