Abstract
A desing of in inductively coupled coil for transcutaneous energy transmission in the wide range of electrical power from 1 mW to 20 W is described. The implantable coils with the amorphous wires have features such as: thin and flexible shape, low temperature rise of 3 degree in the case of the transmission of 20 watts. The coils were applied for artificial heart system and functional electrical stimulation system.
Original language | English |
---|---|
Pages (from-to) | 1276-1282 |
Number of pages | 7 |
Journal | IEEE Transactions on Magnetics |
Volume | 31 |
Issue number | 2 |
DOIs | |