Handling the polarization of soft X-ray using transmission multilayers. A case study

T. Hatano, W. Hu, M. Yamamoto, M. Watanabe

Research output: Contribution to journalArticlepeer-review

6 Citations (Scopus)

Abstract

At the beamline BL8B1, UVSOR, the polarization diagnosis was performed with an ellipsometric method using a so-called rotating analyzer mounted with a reflection multilayer. The performance of a free-standing Al/YB6 transmission multilayer was studied using linearly and circularly polarized light obtained by selecting the angle of observation. The polarizance and the phase retardation of the multilayer for monochromatized light at 65 eV were 92% and 68° at their maxima in the angle of incidence response, respectively. It has been shown that the polarization of synchrotron radiation in the soft X-ray region can be transformed into linear and circular polarization by use of the transmission multilayers.

Original languageEnglish
Pages (from-to)311-314
Number of pages4
JournalJournal of Electron Spectroscopy and Related Phenomena
Volume92
Issue number1-3
DOIs
Publication statusPublished - 1998 May

Keywords

  • Phase retarder
  • Polarizer
  • Synchrotron radiation
  • Transmission multilayer

Fingerprint

Dive into the research topics of 'Handling the polarization of soft X-ray using transmission multilayers. A case study'. Together they form a unique fingerprint.

Cite this