T1 - Hybrid operating room system for the treatment of thoracic and abdominal aortic aneurysms
T2 - Evaluation of the radiation dose received by patients
AU - Haga, Yoshihiro
AU - Chida, Koichi
AU - Sota, Masahiro
AU - Kaga, Yuji
AU - Abe, Mitsuya
AU - Inaba, Yohei
AU - Suzuki, Masatoshi
AU - Meguro, Taiichiro
AU - Zuguchi, Masayuki
N1 - Publisher Copyright:
© 2020 by the authors. Licensee MDPI, Basel, Switzerland.
PY - 2020/10/19
Y1 - 2020/10/19
N2 - In recent years, endovascular treatment of aortic aneurysms has attracted considerable attention as a promising alternative to traditional surgery. Hybrid operating room systems (HORSs) are increasingly being used to perform endovascular procedures. The clinical benefits of endovascular treatments using HORSs are very clear, and these procedures are increasing in number. In procedures such as thoracic endovascular aortic repair (TEVAR) and endovascular aortic repair (EVAR), wires and catheters are used to deliver and deploy the stent graft in the thoracic/abdominal aorta under fluoroscopic control, including DSA. Thus, the radiation dose to the patient is an important issue. We determined radiation dose indicators (the dose–area product (DAP) and air karma (AK) parameters) associated with endovascular treatments (EVAR and TEVAR) using a HORS. As a result, the mean ± standard deviation (SD) DAPs of TEVAR and EVAR were 323.7 ± 161.0 and 371.3 ± 186.0 Gy x cm2, respectively. The mean ± SD AKs of TEVAR and EVAR were 0.92 ± 0.44 and 1.11 ± 0.54 Gy, respectively. The mean ± SD fluoroscopy times of TEVAR and EVAR were 13.4 ± 7.1 and 23.2 ± 11.7 min, respectively. Patient radiation dose results in this study of endovascular treatments using HORSs showed no deterministic radiation effects, such as skin injuries. However, radiation exposure during TEVAR and EVAR cannot be ignored. The radiation dose should be evaluated in HORSs during endovascular treatments. Reducing/optimizing the radiation dose to the patient in HORSs is important.
AB - In recent years, endovascular treatment of aortic aneurysms has attracted considerable attention as a promising alternative to traditional surgery. Hybrid operating room systems (HORSs) are increasingly being used to perform endovascular procedures. The clinical benefits of endovascular treatments using HORSs are very clear, and these procedures are increasing in number. In procedures such as thoracic endovascular aortic repair (TEVAR) and endovascular aortic repair (EVAR), wires and catheters are used to deliver and deploy the stent graft in the thoracic/abdominal aorta under fluoroscopic control, including DSA. Thus, the radiation dose to the patient is an important issue. We determined radiation dose indicators (the dose–area product (DAP) and air karma (AK) parameters) associated with endovascular treatments (EVAR and TEVAR) using a HORS. As a result, the mean ± standard deviation (SD) DAPs of TEVAR and EVAR were 323.7 ± 161.0 and 371.3 ± 186.0 Gy x cm2, respectively. The mean ± SD AKs of TEVAR and EVAR were 0.92 ± 0.44 and 1.11 ± 0.54 Gy, respectively. The mean ± SD fluoroscopy times of TEVAR and EVAR were 13.4 ± 7.1 and 23.2 ± 11.7 min, respectively. Patient radiation dose results in this study of endovascular treatments using HORSs showed no deterministic radiation effects, such as skin injuries. However, radiation exposure during TEVAR and EVAR cannot be ignored. The radiation dose should be evaluated in HORSs during endovascular treatments. Reducing/optimizing the radiation dose to the patient in HORSs is important.
KW - Aneurysm
KW - Disaster medicine
KW - Endovascular treatment
KW - Fluoroscopy
KW - Interventional radiology (IVR)
KW - Radiation dose
KW - Radiation safety
KW - Stent graft
KW - X-ray examination
UR - http://www.scopus.com/inward/record.url?scp=85093510872&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85093510872&partnerID=8YFLogxK
U2 - 10.3390/diagnostics10100846
DO - 10.3390/diagnostics10100846
M3 - Article
AN - SCOPUS:85093510872
SN - 2075-4418
VL - 10
JO - Diagnostics
JF - Diagnostics
IS - 10
M1 - 846
ER -