TY - JOUR
T1 - Multimodality non-rigid image registration for planning, targeting and monitoring during CT-guided percutaneous liver tumor cryoablation
AU - Elhawary, Haytham
AU - Oguro, Sota
AU - Tuncali, Kemal
AU - Morrison, Paul R.
AU - Tatli, Servet
AU - Shyn, Paul B.
AU - Silverman, Stuart G.
AU - Hata, Nobuhiko
N1 - Funding Information:
This publication was made possible by grants 5U41RR019703 , 1R01CA124377 , and 5U54EB005149 from the National Institutes of Health (Bethesda, MD) . Its contents are solely the responsibility of the authors and do not necessarily represent the official views of the National Institutes of Health. Part of this study was funded by the Intelligent Surgical Instruments Project of METI (Japan) .
PY - 2010/11
Y1 - 2010/11
N2 - Rationale and Objectives: The aim of this study was to develop non-rigid image registration between preprocedure contrast-enhanced magnetic resonance (MR) images and intraprocedure unenhanced computed tomographic (CT) images, to enhance tumor visualization and localization during CT imaging-guided liver tumor cryoablation procedures. Materials and Methods: A non-rigid registration technique was evaluated with different preprocessing steps and algorithm parameters and compared to a standard rigid registration approach. The Dice similarity coefficient, target registration error, 95th-percentile Hausdorff distance, and total registration time (minutes) were compared using a two-sided Student's t test. The entire registration method was then applied during five CT imaging-guided liver cryoablation cases with the intraprocedural CT data transmitted directly from the CT scanner, with both accuracy and registration time evaluated. Results: Selected optimal parameters for registration were a section thickness of 5 mm, cropping the field of view to 66% of its original size, manual segmentation of the liver, B-spline control grid of 5 × 5 × 5, and spatial sampling of 50,000 pixels. A mean 95th-percentile Hausdorff distance of 3.3 mm (a 2.5 times improvement compared to rigid registration, P < .05), a mean Dice similarity coefficient of 0.97 (a 13% increase), and a mean target registration error of 4.1 mm (a 2.7 times reduction) were measured. During the cryoablation procedure, registration between the preprocedure MR and the planning intraprocedure CT imaging took a mean time of 10.6 minutes, MR to targeting CT image took 4 minutes, and MR to monitoring CT imaging took 4.3 minutes. Mean registration accuracy was <3.4 mm. Conclusions: Non-rigid registration allowed improved visualization of the tumor during interventional planning, targeting, and evaluation of tumor coverage by the ice ball. Future work is focused on reducing segmentation time to make the method more clinically acceptable.
AB - Rationale and Objectives: The aim of this study was to develop non-rigid image registration between preprocedure contrast-enhanced magnetic resonance (MR) images and intraprocedure unenhanced computed tomographic (CT) images, to enhance tumor visualization and localization during CT imaging-guided liver tumor cryoablation procedures. Materials and Methods: A non-rigid registration technique was evaluated with different preprocessing steps and algorithm parameters and compared to a standard rigid registration approach. The Dice similarity coefficient, target registration error, 95th-percentile Hausdorff distance, and total registration time (minutes) were compared using a two-sided Student's t test. The entire registration method was then applied during five CT imaging-guided liver cryoablation cases with the intraprocedural CT data transmitted directly from the CT scanner, with both accuracy and registration time evaluated. Results: Selected optimal parameters for registration were a section thickness of 5 mm, cropping the field of view to 66% of its original size, manual segmentation of the liver, B-spline control grid of 5 × 5 × 5, and spatial sampling of 50,000 pixels. A mean 95th-percentile Hausdorff distance of 3.3 mm (a 2.5 times improvement compared to rigid registration, P < .05), a mean Dice similarity coefficient of 0.97 (a 13% increase), and a mean target registration error of 4.1 mm (a 2.7 times reduction) were measured. During the cryoablation procedure, registration between the preprocedure MR and the planning intraprocedure CT imaging took a mean time of 10.6 minutes, MR to targeting CT image took 4 minutes, and MR to monitoring CT imaging took 4.3 minutes. Mean registration accuracy was <3.4 mm. Conclusions: Non-rigid registration allowed improved visualization of the tumor during interventional planning, targeting, and evaluation of tumor coverage by the ice ball. Future work is focused on reducing segmentation time to make the method more clinically acceptable.
KW - B-spline registration
KW - Liver tumor cryoablation
KW - Multimodal registration
KW - Non-rigid registration
UR - http://www.scopus.com/inward/record.url?scp=77957683622&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=77957683622&partnerID=8YFLogxK
U2 - 10.1016/j.acra.2010.06.004
DO - 10.1016/j.acra.2010.06.004
M3 - Article
C2 - 20817574
AN - SCOPUS:77957683622
SN - 1076-6332
VL - 17
SP - 1334
EP - 1344
JO - Academic Radiology
JF - Academic Radiology
IS - 11
ER -