N2 - For the development of the new therapeutic device for the atrial fibrillation, implantable cooling device using Peltier element was developed in this study. An implantable cooling device had been consisted from Peltier element with transcutaneous energy transmission system (TETS). 1st coil can be contacted from outside of the body, when the patients will feel palpitation. Electrical current will be induced to the implanted 2nd coil. Peltier element will able to cool the surface of the atrium. For the confirmation of the effect of the cooling device, trial manufacture model was developed. Animal experiments using six healthy adult goats after animal ethical committee allowance was carried out. Fourth intercostals space had been opened after anesthesia inhalation, and various sensors had been inserted. AF was induced by the electrical current with battery. As the results, AF had been recovered to the normal sinus rhythm after cooling in all six goats. So, this cooling system for the control of AF showed evident effect in these experiments. Smaller size cooling device has been under development aiming at totally implantable type. Catheter type cooling device for the insertion by the use of fiber-scope type is now under planning for the clinical application. This new type device may be able to become good news for the patients with uncontrollable AF
AB - For the development of the new therapeutic device for the atrial fibrillation, implantable cooling device using Peltier element was developed in this study. An implantable cooling device had been consisted from Peltier element with transcutaneous energy transmission system (TETS). 1st coil can be contacted from outside of the body, when the patients will feel palpitation. Electrical current will be induced to the implanted 2nd coil. Peltier element will able to cool the surface of the atrium. For the confirmation of the effect of the cooling device, trial manufacture model was developed. Animal experiments using six healthy adult goats after animal ethical committee allowance was carried out. Fourth intercostals space had been opened after anesthesia inhalation, and various sensors had been inserted. AF was induced by the electrical current with battery. As the results, AF had been recovered to the normal sinus rhythm after cooling in all six goats. So, this cooling system for the control of AF showed evident effect in these experiments. Smaller size cooling device has been under development aiming at totally implantable type. Catheter type cooling device for the insertion by the use of fiber-scope type is now under planning for the clinical application. This new type device may be able to become good news for the patients with uncontrollable AF
UR - http://www.scopus.com/inward/record.url?scp=84880842053&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84880842053&partnerID=8YFLogxK
U2 - 10.1109/EMBC.2012.6347298
DO - 10.1109/EMBC.2012.6347298
M3 - Conference contribution
C2 - 23367233
AN - SCOPUS:84880842053
SN - 9781424441198
T3 - Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS
SP - 5741
EP - 5744
BT - 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2012
T2 - 34th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS 2012
Y2 - 28 August 2012 through 1 September 2012
ER -