TY - JOUR
T1 - Precise practical measurements of 90Y used in radiopharmaceuticals using re-entrant ionization chambers
AU - Yamada, Takahiro
AU - Ishii, Keizo
N1 - Copyright:
Copyright 2013 Elsevier B.V., All rights reserved.
PY - 2012/9
Y1 - 2012/9
N2 - For diagnostic nuclear medicine, radiopharmaceuticals are labeled with γ-ray emitting nuclides. Routine activity measurements for these drugs are mainly performed with re-entrant ionization chambers. For radiotherapeutic applications, Β-ray emitting nuclides are also used. Since very few gamma transitions are involved in these nuclides, an ionization chamber responds only to weak bremsstrahlung. In order to apply the re-entrant chamber method to routine activity assay for these Β-ray emitting nuclides, detailed studies were carried out. It was determined that the response was strongly affected by the choice of thickness of the inner wall of the well in the measurement of high energy beta particle emitting nuclides. The geometrical dependences were successfully reduced. Accordingly, routine assay of radioactivity of 90Y therapeutic pharmaceuticals can be performed within an acceptable uncertainty with ordinary re-entrant dose calibrators that are commercially available and equipped in most hospitals.
AB - For diagnostic nuclear medicine, radiopharmaceuticals are labeled with γ-ray emitting nuclides. Routine activity measurements for these drugs are mainly performed with re-entrant ionization chambers. For radiotherapeutic applications, Β-ray emitting nuclides are also used. Since very few gamma transitions are involved in these nuclides, an ionization chamber responds only to weak bremsstrahlung. In order to apply the re-entrant chamber method to routine activity assay for these Β-ray emitting nuclides, detailed studies were carried out. It was determined that the response was strongly affected by the choice of thickness of the inner wall of the well in the measurement of high energy beta particle emitting nuclides. The geometrical dependences were successfully reduced. Accordingly, routine assay of radioactivity of 90Y therapeutic pharmaceuticals can be performed within an acceptable uncertainty with ordinary re-entrant dose calibrators that are commercially available and equipped in most hospitals.
KW - Bremsstrahlung
KW - Radioactivity
KW - Radionuclide calibrators
KW - Radiopharmaceuticals
KW - Re-entrant ionization chamber
KW - Y
UR - http://www.scopus.com/inward/record.url?scp=84864115521&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84864115521&partnerID=8YFLogxK
U2 - 10.1016/j.apradiso.2012.02.096
DO - 10.1016/j.apradiso.2012.02.096
M3 - Article
C2 - 22424744
AN - SCOPUS:84864115521
SN - 0969-8043
VL - 70
SP - 2246
EP - 2250
JO - Applied Radiation and Isotopes
JF - Applied Radiation and Isotopes
IS - 9
ER -