TY - JOUR
T1 - Radiation dose analysis in interventional neuroradiology of unruptured aneurysm cases
AU - Kawauchi, Satoru
AU - Chida, Koichi
AU - Moritake, Takashi
AU - Hamada, Yusuke
AU - Tsuruta, Wataro
N1 - Publisher Copyright:
© 2024 Society for Radiological Protection. Published on behalf of SRP by IOP Publishing Limited. All rights, including for text and data mining, AI training, and similar technologies, are reserved.
PY - 2024/9/1
Y1 - 2024/9/1
N2 - This study aimed to evaluate the radiation doses (peak skin dose (PSD) and bilateral lens dose) for each interventional neuroradiology procedure. A direct measurement system consisting of small radiophotoluminescence glass dosimeter chips and a dosimetry cap made of thin stretchable polyester was used for radiation dosimetry. The mean PSDs for each procedure were 1565 ± 590 mGy (simple technique coil embolization (STCE) cases), 1851 ± 825 mGy (balloon-assisted coil embolization (BACE) cases), 2583 ± 967 mGy (stent-assisted coil embolization (SACE) cases), 1690 ± 597 mGy (simple flow-diverter stenting (FDS) cases), and 2214 ± 726 mGy (FDS + coiling cases). The mean PSD was higher in SACE cases than in STCE, BACE, and simple FDS cases. Moreover, the PSD exceeded 2000 mGy and 3000 mGy in 46 (67.6%) and 19 (27.9%) SACE cases, respectively. The left lens doses for each procedure were 126 ± 111 mGy (STCE cases), 163 ± 152 mGy (BACE cases), 184 ± 148 mGy (SACE cases), 144 ± 60 mGy (simple FDS cases), and 242 ± 178 mGy (FDS + coiling cases). The left lens dose in SACE cases was higher than that in STCE cases and exceeded 500 mGy in 3 (4.4%) patients. In FDS + coiling cases, the mean PSD and left lens dose were 2214 ± 726 mGy and 242 ± 178 mGy, respectively. The left lens dose was higher than that in the STCE and BACE cases, with two (15.4%) patients receiving doses >500 mGy in FDS + coiling cases. The detailed data obtained in this study are expected to contribute to the promotion of radiation dose optimization.
AB - This study aimed to evaluate the radiation doses (peak skin dose (PSD) and bilateral lens dose) for each interventional neuroradiology procedure. A direct measurement system consisting of small radiophotoluminescence glass dosimeter chips and a dosimetry cap made of thin stretchable polyester was used for radiation dosimetry. The mean PSDs for each procedure were 1565 ± 590 mGy (simple technique coil embolization (STCE) cases), 1851 ± 825 mGy (balloon-assisted coil embolization (BACE) cases), 2583 ± 967 mGy (stent-assisted coil embolization (SACE) cases), 1690 ± 597 mGy (simple flow-diverter stenting (FDS) cases), and 2214 ± 726 mGy (FDS + coiling cases). The mean PSD was higher in SACE cases than in STCE, BACE, and simple FDS cases. Moreover, the PSD exceeded 2000 mGy and 3000 mGy in 46 (67.6%) and 19 (27.9%) SACE cases, respectively. The left lens doses for each procedure were 126 ± 111 mGy (STCE cases), 163 ± 152 mGy (BACE cases), 184 ± 148 mGy (SACE cases), 144 ± 60 mGy (simple FDS cases), and 242 ± 178 mGy (FDS + coiling cases). The left lens dose in SACE cases was higher than that in STCE cases and exceeded 500 mGy in 3 (4.4%) patients. In FDS + coiling cases, the mean PSD and left lens dose were 2214 ± 726 mGy and 242 ± 178 mGy, respectively. The left lens dose was higher than that in the STCE and BACE cases, with two (15.4%) patients receiving doses >500 mGy in FDS + coiling cases. The detailed data obtained in this study are expected to contribute to the promotion of radiation dose optimization.
KW - lens dose
KW - neurointerventional radiology
KW - peak skin dose
KW - radiophotoluminescent glass dosimeter
KW - unruptured aneurysm
UR - http://www.scopus.com/inward/record.url?scp=85203878856&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85203878856&partnerID=8YFLogxK
U2 - 10.1088/1361-6498/ad76b3
DO - 10.1088/1361-6498/ad76b3
M3 - Article
C2 - 39226910
AN - SCOPUS:85203878856
SN - 0952-4746
VL - 44
JO - Journal of Radiological Protection
JF - Journal of Radiological Protection
IS - 3
M1 - 031515
ER -