Abstract
A selective off-line preconcentration technique for the V(V) complex with 4-(2-pyridylazo)resorcinol has been developed and successfully applied to the determination of V(V) in an air-borne sample. The target complex was separated from excess reagent using an ODS cartridge and water as the eluent. The complex was then concentrated on another ODS cartridge using tetrabutylammonium bromide and eluted with methanol; the eluate was applied to a one-drop concentration/HPLC. A detection limit as low as (6.05 ± 0.82)× 10-11 M (5 ppt) was achieved. 2006
Original language | English |
---|---|
Pages (from-to) | 1585-1588 |
Number of pages | 4 |
Journal | analytical sciences |
Volume | 22 |
Issue number | 12 |
DOIs | |
Publication status | Published - 2006 Dec |
Externally published | Yes |
ASJC Scopus subject areas
- Analytical Chemistry