Abstract
The energy resolutions of 976keV conversion electrons from a 207Bi source are measured in a gridded ionization chamber filled with tetramethylsilane(TMS) and tetramethylgermanium(TMG), and are found to be about 5.71, 5.53%(rms) for TMS and TMG, respectively. We discuss on the difference between the obtained energy resolution and the theoretically limited energy resolution. We also deduce a simple method of estimating the electron lifetime by a gridded ionization chamber. The electron lifetime, free ion yield and thermalization length for these liquids are measured by this simple method.
Original language | English |
---|---|
Pages | 705-709 |
Number of pages | 5 |
Publication status | Published - 1995 |
Event | Proceedings of the 1995 IEEE Nuclear Science Symposium and Medical Imaging Conference. Part 1 (of 3) - San Francisco, CA, USA Duration: 1995 Oct 21 → 1995 Oct 28 |
Conference
Conference | Proceedings of the 1995 IEEE Nuclear Science Symposium and Medical Imaging Conference. Part 1 (of 3) |
---|---|
City | San Francisco, CA, USA |
Period | 95/10/21 → 95/10/28 |