The test measurements of TOF in small and large scintillation counters

T. Iwata, K. Dohmaé, S. Fukui, C. Y. Gao, I. Hirose, N. Horikawa, T. Nakanishi, S. Sakanaka, T. Toyama, T. Kitagaki, Y. Ohtani, H. Sagawa, Akira Yamaguchi, M. Mutoh, R. Yamamoto

Research output: Contribution to journalArticlepeer-review

1 Citation (Scopus)

Abstract

We tested the time resolution of "small" scintillation counters (0.8 (height) × 2.6 (width) cm2 and 0.7 (height) × 2.1 (width) cm2, with 1.0 cm thickness) and of a "large" counter (21 (height) × 96 (width) cm2 with 2 cm thickness). The time walk correction using software (off-line) was compared with that using hardware (constant fraction discriminator). It was found that the two methods gave nearly equal time resolution. Intrinsic time resolutions of 155 and 370 ps fwhm were obtained for the small and the large counters, respectively.

Original languageEnglish
Pages (from-to)434-439
Number of pages6
JournalNuclear Inst. and Methods in Physics Research, A
Volume244
Issue number3
DOIs
Publication statusPublished - 1986 Apr 1

ASJC Scopus subject areas

  • Nuclear and High Energy Physics
  • Instrumentation