TY - JOUR
T1 - Topography, microhardness, and precision of fit on ready-made zirconia abutment before/after sintering process
AU - Kanno, Taro
AU - Milleding, Percy
AU - Wennerberg, Ann
PY - 2007/9
Y1 - 2007/9
N2 - Background Sintering porcelain on a ceramic abutment may change the microstructure and result in aging processes that influence the mechanical properties, internal strain, and the three-dimensional form of the abutment, thus causing a possible misfit between the abutment and the fixture. Purpose The aim was to investigate topography, microhardness, and precision of fit on yttrium-stabilized zirconia (Y-TZP) abutments before/after the sintering process. Materials and Methods Ten Y-TZP abutment samples were ground to a shape used in the clinical situation and divided at random into two groups: before/after sintering. After the surface roughness was measured on all abutments, the abutments were connected to fixture replicas, embedded in resin, and cut in the longitudinal axis. Both sides of the cut samples were measured with respect to microhardness and minimum distance between fixture and abutment surface. t-Test, one-way analysis of variance, and Bonferroni multiple comparisons were used to investigate statistical significant differences. Results The surface roughness (Sa and Sdr) after sintering was significantly higher than before sintering. The total average values of microhardness after sintering were statistically lower than before sintering with a difference of 2. The total distance between abutment/fixture before/after sintering demonstrated no statistically significant difference. Contact between abutment/fixture was most common at the top area of the fixture. Conclusion A slight decrease of microhardness and contamination of porcelain particles immediately below the veneered part were found on the Y-TZP abutment after sintering. The sintering process did not affect the precision of fit.
AB - Background Sintering porcelain on a ceramic abutment may change the microstructure and result in aging processes that influence the mechanical properties, internal strain, and the three-dimensional form of the abutment, thus causing a possible misfit between the abutment and the fixture. Purpose The aim was to investigate topography, microhardness, and precision of fit on yttrium-stabilized zirconia (Y-TZP) abutments before/after the sintering process. Materials and Methods Ten Y-TZP abutment samples were ground to a shape used in the clinical situation and divided at random into two groups: before/after sintering. After the surface roughness was measured on all abutments, the abutments were connected to fixture replicas, embedded in resin, and cut in the longitudinal axis. Both sides of the cut samples were measured with respect to microhardness and minimum distance between fixture and abutment surface. t-Test, one-way analysis of variance, and Bonferroni multiple comparisons were used to investigate statistical significant differences. Results The surface roughness (Sa and Sdr) after sintering was significantly higher than before sintering. The total average values of microhardness after sintering were statistically lower than before sintering with a difference of 2. The total distance between abutment/fixture before/after sintering demonstrated no statistically significant difference. Contact between abutment/fixture was most common at the top area of the fixture. Conclusion A slight decrease of microhardness and contamination of porcelain particles immediately below the veneered part were found on the Y-TZP abutment after sintering. The sintering process did not affect the precision of fit.
KW - Ceramic abutment
KW - Dental implant
KW - Mechanical properties
KW - Precision of fit
KW - Surface roughness
KW - Y-TZP
UR - http://www.scopus.com/inward/record.url?scp=34548067979&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=34548067979&partnerID=8YFLogxK
U2 - 10.1111/j.1708-8208.2007.00044.x
DO - 10.1111/j.1708-8208.2007.00044.x
M3 - Article
C2 - 17716260
AN - SCOPUS:34548067979
SN - 1523-0899
VL - 9
SP - 156
EP - 165
JO - Clinical Implant Dentistry and Related Research
JF - Clinical Implant Dentistry and Related Research
IS - 3
ER -