Ultrasonic measurement of arterial pressure and diameter at the same position

Kota Kudo, Hiroshi Kanai, Kazuto Kobayashi

Research output: Contribution to journalArticlepeer-review

Abstract

For early detection of disorder in the vascular endothelial function, a technique to evaluate the viscoelastic characteristic of the vascular wall was developed by measuring the blood pressure of the radial artery and measuring the change of the vessel diameter at a different close position. By it measurement, however, an accurate evaluation of the viscoelastic characteristic is difficult due to the small time lag between the waveforms. Therefore, for measuring both the blood pressure and the change of the vessel diameter at the same position, the present study presents a new method to acquire the blood pressure by the same ultrasonic probe as that for the ultrasonic measurement of the change of the vessel diameter. From the resultant waveforms, the hysteresis characteristic of the vessel diameter was obtained in in vivo experiments.

Original languageEnglish
Pages (from-to)599-600
Number of pages2
JournalTransactions of Japanese Society for Medical and Biological Engineering
Volume55
Issue numberProc
DOIs
Publication statusPublished - 2017

Keywords

  • Biological engineering
  • Hysteresis characteristic
  • Radial artery
  • Ultrasound

Fingerprint

Dive into the research topics of 'Ultrasonic measurement of arterial pressure and diameter at the same position'. Together they form a unique fingerprint.

Cite this