TY - JOUR
T1 - Whole gamma imaging
T2 - A new concept of PET combined with Compton imaging
AU - Yoshida, Eiji
AU - Tashima, Hideaki
AU - Nagatsu, Kotaro
AU - Tsuji, Atsushi B.
AU - Kamada, Kei
AU - Parodi, Katia
AU - Yamaya, Taiga
N1 - Publisher Copyright:
© 2020 Institute of Physics and Engineering in Medicine.
PY - 2020/6/21
Y1 - 2020/6/21
N2 - We proposed a concept of whole gamma imaging (WGI) that utilizes all detectable gamma rays for imaging. An additional detector ring, which is used as the scatterer, is inserted in the field-of-view of a PET ring so that single gamma rays can be detected by the Compton imaging method. In particular, for the non-pure positron emitters which emit an additional gamma ray almost at the same time, triple gamma imaging will be enabled; localization on each line-of-response (LOR) is possible by using the Compton cone of the additional gamma ray. We developed a prototype to show a proof of the WGI concept. The diameters of scatterer ring and PET ring were set as 20 cm and 66 cm, respectively. For Compton imaging of the 662-keV gamma ray from a 137Cs point source, spatial resolution obtained by the list-mode OSEM algorithm was 4.4 mm FWHM at the 8 cm off-center position and 13.1 mm FWHM at the center position. For PET imaging of a 22Na point source, spatial resolution was about 2 mm FWHM at all positions. For the triple gamma imaging, 5.7 mm FWHM (center) and 4.8 mm FWHM (8 cm off-center) were obtained for the 22Na point source just by plotting the intersecting points between each LOR and each Compton cone of the 1275-keV gamma ray. No image reconstruction was applied. Scandium-44 was produced as a practical candidate of the non-pure positron emitters, and 6.6 mm FWHM (center) and 5.8 mm FWHM (8 cm off-center) were obtained in the same manner. This direct imaging approach which neither requires time-consuming event integration nor iterative image reconstruction may allow in vivo real-Time tracking of a tiny amount of activity. Our initial results showed the feasibility of the WGI concept, which is a novel combination of PET and Compton imaging.
AB - We proposed a concept of whole gamma imaging (WGI) that utilizes all detectable gamma rays for imaging. An additional detector ring, which is used as the scatterer, is inserted in the field-of-view of a PET ring so that single gamma rays can be detected by the Compton imaging method. In particular, for the non-pure positron emitters which emit an additional gamma ray almost at the same time, triple gamma imaging will be enabled; localization on each line-of-response (LOR) is possible by using the Compton cone of the additional gamma ray. We developed a prototype to show a proof of the WGI concept. The diameters of scatterer ring and PET ring were set as 20 cm and 66 cm, respectively. For Compton imaging of the 662-keV gamma ray from a 137Cs point source, spatial resolution obtained by the list-mode OSEM algorithm was 4.4 mm FWHM at the 8 cm off-center position and 13.1 mm FWHM at the center position. For PET imaging of a 22Na point source, spatial resolution was about 2 mm FWHM at all positions. For the triple gamma imaging, 5.7 mm FWHM (center) and 4.8 mm FWHM (8 cm off-center) were obtained for the 22Na point source just by plotting the intersecting points between each LOR and each Compton cone of the 1275-keV gamma ray. No image reconstruction was applied. Scandium-44 was produced as a practical candidate of the non-pure positron emitters, and 6.6 mm FWHM (center) and 5.8 mm FWHM (8 cm off-center) were obtained in the same manner. This direct imaging approach which neither requires time-consuming event integration nor iterative image reconstruction may allow in vivo real-Time tracking of a tiny amount of activity. Our initial results showed the feasibility of the WGI concept, which is a novel combination of PET and Compton imaging.
KW - Compton camera
KW - PET
KW - triple gamma imaging
UR - http://www.scopus.com/inward/record.url?scp=85087158046&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85087158046&partnerID=8YFLogxK
U2 - 10.1088/1361-6560/ab8e89
DO - 10.1088/1361-6560/ab8e89
M3 - Article
C2 - 32348968
AN - SCOPUS:85087158046
SN - 0031-9155
VL - 65
JO - Physics in Medicine and Biology
JF - Physics in Medicine and Biology
IS - 12
M1 - 125013
ER -