Advanced scanning electrochemical microscope system for high-resolution imaging and electrochemical applications

Ryo Matsuoka, Shigeo Aoyagi, Naoshi Matsumoto, Masaaki Matsudaira, Yasufumi Takahashi, Akichika Kumatani, Hiroki Ida, Hirokazu Munakata, Katsuhiko Iida, Hitoshi Shiku, Kiyoshi Kanamura, Tomokazu Matsue

研究成果: ジャーナルへの寄稿学術論文査読

2 被引用数 (Scopus)

抄録

We developed three modes of high resolution scanning probe microscope system based on electrochemical principles, scanning ion conductance microscopy (SICM), scanning electrochemical microscopy-scanning ion conductance microscopy (SECM-SICM), and scanning electrochemical cell microscopy (SECCM). Firstly, a developed SICM system was constructed with a nanopipette filled with electrolyte solution as a probe. High resolution topographic images of NanoCulture®Plate with hexagonal chambers with 2 μm-width banks, electrodeposited PEDOT (poly(3,4-ethylenedioxythiophane)) film, and fixed human squamous cell carcinoma were captured by the SICM. Second, SECM-SICM was performed with a double-barrel carbon nanoprobe. The one barrel of the nanoprobe was filled with carbon and used for SECM apparatus, the other barrel was filled with electrolyte for SICM configurations. Using the SECM-SICM system, simultaneous topographic and electrochemical images of micro-band electrodes with 10 μm-width line and space, and a cathode site of corrosion in the aluminum die gusto with submicron spatial resolution were obtained. Thirdly, the SECCM featuring a nanopipette probe filled with LiCl electrolyte was applied for obtaining topographies and images of current activity of a LiFePO4 electrode.

本文言語英語
ページ(範囲)319-326
ページ数8
ジャーナルElectrochemistry
85
6
DOI
出版ステータス出版済み - 2017

フィンガープリント

「Advanced scanning electrochemical microscope system for high-resolution imaging and electrochemical applications」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル