TY - JOUR
T1 - Comparison of the intubating laryngeal mask airway and laryngeal tube placement during manual in-line stabilisation of the neck
AU - Komatsu, R.
AU - Nagata, O.
AU - Kamata, K.
AU - Yamagata, K.
AU - Sessler, D. I.
AU - Ozaki, M.
PY - 2005/2
Y1 - 2005/2
N2 - We compared the placement of the laryngeal tube (LT) with the intubating laryngeal mask airway (ILMA) in 51 patients whose necks were stabilised by manual in-line traction. Following induction of anaesthesia and neuromuscular blockade, the LT and ILMA were inserted consecutively in a randomised, crossover design. Using pressure-controlled ventilation (20 cmH2O inspiratory pressure), we measured insertion attempts, time to establish positive-pressure ventilation, tidal volume, gastric insufflation, and minimum airway pressure at which gas leaked around the cuff. Data were compared using Wilcoxon signed-rank tests; p < 0.05 was considered significant. Insertion was found to be more difficult with the LT (successful at first attempt in 16 patients) than with the ILMA (successful at first attempt in 42 patients, p < 0.0001). Time required for insertion was longer for the LT (28 [23-35] s, median [interquartile range]) than for the ILMA (20 [15-25] s, p = 0.0009). Tidal volume was less for the LT (440 [290-670] ml) than for the ILMA (630 [440-750] ml, p = 0.013). Minimum airway pressure at which gas leak occurred and incidence of gastric insufflation were similar with two devices. In patients whose necks were stabilised with manual in-line traction, insertion of the ILMA was easier and quicker than insertion of the LT and tidal volume was greater with the ILMA than the LT.
AB - We compared the placement of the laryngeal tube (LT) with the intubating laryngeal mask airway (ILMA) in 51 patients whose necks were stabilised by manual in-line traction. Following induction of anaesthesia and neuromuscular blockade, the LT and ILMA were inserted consecutively in a randomised, crossover design. Using pressure-controlled ventilation (20 cmH2O inspiratory pressure), we measured insertion attempts, time to establish positive-pressure ventilation, tidal volume, gastric insufflation, and minimum airway pressure at which gas leaked around the cuff. Data were compared using Wilcoxon signed-rank tests; p < 0.05 was considered significant. Insertion was found to be more difficult with the LT (successful at first attempt in 16 patients) than with the ILMA (successful at first attempt in 42 patients, p < 0.0001). Time required for insertion was longer for the LT (28 [23-35] s, median [interquartile range]) than for the ILMA (20 [15-25] s, p = 0.0009). Tidal volume was less for the LT (440 [290-670] ml) than for the ILMA (630 [440-750] ml, p = 0.013). Minimum airway pressure at which gas leak occurred and incidence of gastric insufflation were similar with two devices. In patients whose necks were stabilised with manual in-line traction, insertion of the ILMA was easier and quicker than insertion of the LT and tidal volume was greater with the ILMA than the LT.
UR - http://www.scopus.com/inward/record.url?scp=13544271084&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=13544271084&partnerID=8YFLogxK
U2 - 10.1111/j.1365-2044.2004.03958.x
DO - 10.1111/j.1365-2044.2004.03958.x
M3 - Article
C2 - 15644005
AN - SCOPUS:13544271084
SN - 0003-2409
VL - 60
SP - 113
EP - 117
JO - Anaesthesia
JF - Anaesthesia
IS - 2
ER -