Dose evaluation for foetal computed tomography with a 320-rowunit in wide-volume mode andan 80-row unit in helical scanning mode: A phantom study

Yuta Matsunaga, Ai Kawaguchi, Masanao Kobayashi, Shigetaka Suzuki, Shoichi Suzuki, Koichi Chida

研究成果: Article査読

10 被引用数 (Scopus)

抄録

The purpose of this study was to evaluate the maternal and foetal effective doses during foetal computed tomography (CT) and to compare the radiation dose, dose profile and image noise on 80-row CT in helical scanning mode and 320-row CT in widevolume scanning mode. The radiation doses were measured using thermoluminescent dosemeters implanted at various organ sites of an anthropomorphic pregnant phantom. The foetal doses in the 320-row multi-detector CT (MDCT) and 80-row MDCT units were higher than the volume CT dose index (CTDIvol). The dose profile in the 320-row MDCToverlapped in two places but showed no overlap in the 80-row MDCT. There were no significant differences in image noise between the two scanning modes. The foetal dose evaluation by CTDIvol may underestimate the foetal radiation risk. When using the wide-volume mode, operators must take into account the number of scans and overlap between volumetric sections.

本文言語English
論文番号ncv374
ページ(範囲)523-530
ページ数8
ジャーナルRadiation protection dosimetry
168
4
DOI
出版ステータスPublished - 2014 12月 8

ASJC Scopus subject areas

  • 放射線
  • 放射線技術および超音波技術
  • 放射線学、核医学およびイメージング
  • 公衆衛生学、環境および労働衛生

フィンガープリント

「Dose evaluation for foetal computed tomography with a 320-rowunit in wide-volume mode andan 80-row unit in helical scanning mode: A phantom study」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル