TY - JOUR
T1 - Feasibility of real-time treatment feedback using novel filter for eliminating therapeutic ultrasound noise with high-speed ultrasonic imaging in ultrasound-guided high-intensity focused ultrasound treatment
AU - Takagi, Ryo
AU - Jimbo, Hayato
AU - Iwasaki, Ryosuke
AU - Tomiyasu, Kentaro
AU - Yoshizawa, Shin
AU - Umemura, Shin Ichiro
N1 - Publisher Copyright:
© 2016 The Japan Society of Applied Physics.
PY - 2016/7
Y1 - 2016/7
N2 - In the conventional ultrasonic monitoring of high-intensity focused ultrasound (HIFU) treatment, a significant interval between HIFU shots is required when monitoring target tissue to avoid interference between HIFU noise and RF echo signals. In our previous study, a new filtering method to eliminate only HIFU noise while maintaining tissue signals intact was proposed, and it was shown that the thermal coagulation could be detected during simultaneous HIFU irradiation through off-line processing. In this study, the filtering method and a real-time coagulation detection algorithm were implemented in an ultrasound imaging system, whose use for sequential exposure with multiple foci was demonstrated similarly to a commercial HIFU ablation system. The coagulation was automatically detected by the proposed method during real-time simultaneous HIFU irradiation, and the HIFU exposure time was controlled according to the changes in the tissue. The results imply that ultrasonic monitoring with the filtering and detection methods is useful for true real-time detection of changes in the tissue due to thermal coagulation during HIFU exposure.
AB - In the conventional ultrasonic monitoring of high-intensity focused ultrasound (HIFU) treatment, a significant interval between HIFU shots is required when monitoring target tissue to avoid interference between HIFU noise and RF echo signals. In our previous study, a new filtering method to eliminate only HIFU noise while maintaining tissue signals intact was proposed, and it was shown that the thermal coagulation could be detected during simultaneous HIFU irradiation through off-line processing. In this study, the filtering method and a real-time coagulation detection algorithm were implemented in an ultrasound imaging system, whose use for sequential exposure with multiple foci was demonstrated similarly to a commercial HIFU ablation system. The coagulation was automatically detected by the proposed method during real-time simultaneous HIFU irradiation, and the HIFU exposure time was controlled according to the changes in the tissue. The results imply that ultrasonic monitoring with the filtering and detection methods is useful for true real-time detection of changes in the tissue due to thermal coagulation during HIFU exposure.
UR - http://www.scopus.com/inward/record.url?scp=84978766742&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84978766742&partnerID=8YFLogxK
U2 - 10.7567/JJAP.55.07KC10
DO - 10.7567/JJAP.55.07KC10
M3 - Article
AN - SCOPUS:84978766742
SN - 0021-4922
VL - 55
JO - Japanese Journal of Applied Physics, Part 1: Regular Papers & Short Notes
JF - Japanese Journal of Applied Physics, Part 1: Regular Papers & Short Notes
IS - 7
M1 - 07KC10
ER -