Impact of selective ion transmission on measurement by retarding field analyzer

H. Takahashi, T. Seino, R. Nishimura, K. Yoshimura, A. Kanno, T. Hara, Y. Takahashi, S. Kagaya, A. Matsuyama, Y. Hayashi, K. Tobita

研究成果: ジャーナルへの寄稿学術論文査読

1 被引用数 (Scopus)

抄録

The impact of selective ion transmission on the use of retarding field analyzers (RFAs), which is attributed to the ion Larmor motion, was investigated. Depending on the ion Larmor radius and ion guiding center position, selective transmission limits the parallel and perpendicular ion velocities inside the RFA volume. The velocity limitation flattens I-V curve, leading to an overestimation of the parallel ion temperature. However, the overestimation can be reduced to an acceptable level using the slope in a large grid potential region. The influence of selective transmission depends on both parallel and perpendicular ion temperatures. This nature likely allows the simultaneous determination of two ion temperatures by the best fit of the experimental data, even from a single I-V curve. Applicability of this method was investigated using a radio frequency plasma.

本文言語英語
論文番号053506
ジャーナルPhysics of Plasmas
30
5
DOI
出版ステータス出版済み - 2023 5月 1

フィンガープリント

「Impact of selective ion transmission on measurement by retarding field analyzer」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル