Large Amplitude Harmonic Vibration under Pulse Doppler Measurement

N. Hermawan, T. Ishii, Y. Yamakoshi, Y. Saijo

研究成果: Conference article査読

抄録

A flow velocity or tissue motion can be measured by pulse Doppler method. However, when the method is applied to a sinusoid motion or harmonic vibration, the estimation result is deviated. In this paper, the generalized theory of pulse Doppler measurement to the sinusoid motion is explained and the relation between the estimation deviation and the real value is given. Furthermore, a special case of binary velocity estimation is expanded and simulated with different parameters. This simulation result suggested that different selection of parameters may fundamentally affect the output of pulse Doppler estimation.

本文言語English
論文番号012034
ジャーナルJournal of Physics: Conference Series
2071
1
DOI
出版ステータスPublished - 2021 11月 25
イベント2021 International Conference on Biomedical Engineering, ICoBE 2021 - Virtual, Online, Malaysia
継続期間: 2021 9月 142021 9月 15

ASJC Scopus subject areas

  • 物理学および天文学(全般)

フィンガープリント

「Large Amplitude Harmonic Vibration under Pulse Doppler Measurement」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル